Medical devices in Great Britain: the MHRA's post-market surveillance rules, and where ISO 13485 fits

Since June 2025 manufacturers of medical devices on the GB market have had stronger post-market surveillance duties, including serious incident reporting within 2, 10 or 15 days. What the MHRA guidance asks for, and how an ISO 13485 quality system supports it.

Published
Reading time
3 min
Sources checked
23 November 2026
The three parts of a hip replacement: a dark metal stem, a ball head and a white plastic cup, on a cream cloth
The parts of a hip replacement: stem, ball head and cup. Photo: Nuno Nogueira (Nmnogueira), Public domain

A medical device's approval for the market is the start of the story, not the end. Once a device is in use, its manufacturer must watch how it performs, spot problems and act on them. In Great Britain those duties were strengthened in 2025. This article summarises the MHRA's guidance and explains how it relates to ISO 13485, the quality management standard written for organisations in the medical device life cycle.

What changed

The MHRA's guidance explains that amending regulations made in 2024 inserted a new Part 4A into the UK Medical Devices Regulations 2002, covering post-market surveillance (PMS), "notification requirements for incidents" and "preventive and corrective actions". The GOV.UK page was updated on 16 June 2025 to reflect them coming into force. The rules cover medical devices in Great Britain, including in vitro diagnostic (IVD) devices and active implantable devices.

The PMS system and plan

According to the MHRA, a manufacturer must define its PMS system in a plan that is "proportionate to the risk posed by the device". At a minimum the plan sets out its objectives, how information is gathered, the "methods of data analysis", how vigilance reporting duties are met, and the links to preventive and corrective action within risk management.

The reports required depend on the device's class:

Device classReportHow often
Class I; IVD class A or BPost-market surveillance report (PMSR)Updated at least every 3 years
Class IIaPeriodic safety update report (PSUR)First within 2 years, then every 2 years
Class IIb, Class III; IVD class C or DPSURFirst within 1 year, then every year

For higher-risk classes, the guidance says the UK approved body reviews the PSUR.

Serious incident reporting

The MHRA sets deadlines that run from when the manufacturer becomes aware of a serious incident:

  • 2 calendar days for a serious public health threat;
  • 10 calendar days for a death or an unanticipated serious deterioration in health;
  • 15 calendar days for other serious incidents.

The guidance is blunt: "Do not delay reporting because of incomplete information" and "If in doubt about whether to report an incident, report it." It also covers field safety corrective actions and field safety notices, and trend reporting — reporting "a significant increase in the number or severity of incidents" against thresholds set in advance.

Where ISO 13485 fits

ISO 13485 sets quality management system requirements for organisations involved in the life cycle of medical devices, written to support regulatory compliance. Its clause 8 already requires feedback processes, complaint handling, reporting to regulatory authorities, control of nonconforming product and corrective and preventive action. So a well-run ISO 13485 system is the natural home for the MHRA's PMS duties:

  • Feedback and complaint handling feed the PMS data the MHRA expects to be analysed.
  • Reporting to regulators is where the 2/10/15-day clock needs to be built into procedures — with clear responsibility for deciding whether an event is reportable.
  • Risk management: the MHRA says PMS data must be an input to the risk management process, which may follow ISO 14971.
  • CAPA links what PMS finds to action.

Remember that ISO 13485 doesn't use the Harmonized Structure; its clause numbers follow the older 2008-era layout, which trips up auditors who know ISO 9001. And certification to ISO 13485 doesn't replace the regulatory requirements — the regulations decide what must be reported and when.

A check for internal auditors

When auditing PMS against the new rules, useful questions include: is there a PMS plan for each device or family, proportionate to its class? Are PSURs or PMSRs on schedule? Could the team show how a recent complaint was assessed for reportability, and how long it took? Are trend thresholds defined? Our guide to writing audit findings helps frame what you find.

Training for ISO 13485